Electrical Specifications

The miniSMU MS01 is designed to provide precise measurement and sourcing capabilities across a wide range of voltage and current values. This section details the electrical specifications of the device.

Voltage Specifications

ParameterSpecificationNotes
Range-12 V to +12 VFull 4-quadrant operation
Accuracy±(0.1% + 1.5 mV)Percentage of reading + offset
Output Impedance< 1 ΩAt DC
Short-Circuit Current180 mALimited by current protection
Settling Time< 10 msTo 0.1% of final value with resistive load

Current Measurement

ParameterSpecificationNotes
Range-180 mA to +180 mAFull span
Accuracy±(0.2% + 10 nA)Percentage of reading + offset
Resolution100 pAIn lowest range

Current Ranges

The miniSMU MS01 features multiple current ranges that can be automatically or manually selected:

RangeSpanResolutionAccuracy
Range 10 to 500 nA100 pA±(0.2% + 10 nA)
Range 2500 nA to 20 μA1 nA±(0.2% + 10 nA)
Range 320 μA to 0.5 mA10 nA±(0.2% + 10 nA)
Range 40.5 mA to 13 mA100 nA±(0.2% + 10 nA)
Range 513 mA to 180 mA1 μA±(0.2% + 10 nA)

Auto-ranging capability allows for optimal precision across the full measurement range.

Current Sourcing

ParameterSpecificationNotes
Range-180 mA to +180 mAFull 4-quadrant operation
Accuracy±(0.2% + 10 nA)Percentage of setting + offset
Resolution100 pAIn lowest range
Setting ResolutionSame as measurement resolutionBy range
Compliance VoltageUp to ±12 VLimit can be set lower
Settling Time< 10 msTo 0.1% of final value with resistive load

Power Specifications

Input Power Options

Power SourceSpecificationMaximum Current Capability
USB-C5VSMU current limited to 50 mA in USB-powered mode by default
DC Barrel12V via DC barrel connectorFull current range available

Power Adapter Specifications

ParameterSpecification
Input90-264 V AC, 50/60 Hz
Output12V 1500 mA
Connector5.5mm outer diameter, 2.1mm inner diameter, center positive
Safety CertificationsUKCA, CE, cULus, TUV, RCM (C-Tick), RoHS, REACH, IEC/EN 62368-1

Power Consumption

ParameterSpecification
Maximum Power Draw15W
Typical Operating Power5W
Standby Power< 1.5W
Off Power< 0.01W
Output PowerMaximum 2.1 W sourced or sunk per channel

Protection Features

Protection TypeSpecification
Overvoltage ProtectionProgrammable from ±0.1V to ±12V
Overcurrent ProtectionProgrammable from ±10 nA to ±180 mA
Thermal ProtectionAutomatic shutdown if internal temperature exceeds safe limits
Input Protection±13V maximum without damage

All specifications are valid under the following conditions:

  • After 30-minute warm-up
  • Ambient temperature 23°C ±5°C
  • Humidity 10-80% RH non-condensing
  • Calibration interval of 1 year

Please note: All specifications are subject to change without notice. Always refer to the most recent documentation for the most up-to-date information.